    Using the bootstrap method to evaluate image noise for investigation of axial collimation in hybrid PET

  • Author

    Groiselle, Corinne J. ; Glick, Stephen J.

  • Author_Institution
    Med. Sch., Univ. of Massachusetts, Worcester, MA, USA
  • Volume
    2
  • fYear
    2002
  • fDate
    10-16 Nov. 2002
  • Firstpage
    782
  • Abstract
    The purpose of this work was to investigate the use of the bootstrap method to evaluate noise in maximum-likelihood, ordered subsets list-mode iterative reconstructions from data simulated on a hybrid PET system. In addition, image noise from reconstructions acquired with various levels of axial collimation was investigated. A warm cylinder (20-cm diameter and 70-cm long) was simulated using the GEANT 3.21 Monte-Carlo simulation code for different collimator designs: open-frame, parallel axial (PAR), and hybrid parallel fan-beam (HPF) axial collimators. Ten noisy sets of list-mode data were simulated for each collimator. In addition, events from the first noisy list-mode data set of each category (open-frame, PAR axial and HPF axial collimators) were chosen at random with replacement to produce ten new sets of the same size. List-mode data were reconstructed using maximum-likelihood, ordered-subsets list-mode iterative reconstruction using 2 iterations and 5 subsets. The total counts within volumes of interest (VOI) were computed, and the variance of this estimate over noise realizations was evaluated. The F-test was used to test statistical significance between variances of VOI estimates from noisy and bootstrap samples. The F-test results indicate that there is no significant difference in the variance of simulations and bootstrap data when using the open-frame, the PAR axial and the HPF axial collimators. The test showed that the precision of estimates vary with the axial spatial location within the object and the axial collimator. In addition, noise variation within list-mode reconstructed images appears to be related to the 3D sensitivity map.
  • Keywords
    Monte Carlo methods; diagnostic radiography; image reconstruction; iterative methods; maximum likelihood estimation; medical computing; noise; positron emission tomography; 3D sensitivity map; Monte-Carlo simulation; axial collimation; bootstrap method; hybrid PET; image noise; iterative reconstructions; maximum-likelihood; noise realizations; statistical significance; Biomedical imaging; Collimators; Computational modeling; Face detection; Head; Image reconstruction; Maximum likelihood estimation; Nuclear medicine; Positron emission tomography; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2002 IEEE
  • Print_ISBN
    0-7803-7636-6
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2002.1239439
  • Filename
    1239439