Title :
Development of external fiducial markers for image registration in small animal SPECT/CT
Author :
Sakdinawat, Anne E. ; Iwata, Koji ; Hwang, Andrew B. ; Tang, Hamilton R. ; Wong, Kenneth H. ; Hasegawa, Bruce H.
Author_Institution :
Dept. of Radiol., California Univ., San Francisco, CA, USA
Abstract :
An image registration technique that can be used to register microCT and microSPECT images acquired with a high-resolution dual-modality SPECT/CT small animal imaging system is being developed. External fiducial markers containing droplets of the K2HPO4 (visible in CT)-99mTc-pertechnetate (visible in SPECT) solution dispersed in narrow-gauge polyethylene tubing were placed on the object prior to imaging. According to phantom measurements, the optimum concentrations of the marker solution were found to be 250 mg/mL for K2HPO4 and approximately 6-8 mCi/mL for 99mTc-pertechnetate. These concentrations were applied for a marker used to register mouse heart phantom images, and were easily visualized in both microSPECT and microCT images.
Keywords :
cardiology; computerised tomography; image registration; medical image processing; phantoms; single photon emission computed tomography; K2HPO4; Tc; external fiducial markers; high-resolution dual-modality; image registration; mouse heart phantom images; phantom measurements; small animal SPECT/CT; Animals; Computed tomography; Heart; High-resolution imaging; Image registration; Imaging phantoms; Mice; Optical imaging; Polyethylene; Visualization;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2002 IEEE
Print_ISBN :
0-7803-7636-6
DOI :
10.1109/NSSMIC.2002.1239456