• DocumentCode
    2196872
  • Title

    First test of liquid Xe scintillation detector for PET

  • Author

    Nishikido, F. ; Doke, T. ; Kikuchi, J. ; Yamamoto, M. ; Mori, T. ; Takizawa, K.

  • Author_Institution
    Waseda Univ., Tokyo, Japan
  • Volume
    3
  • fYear
    2002
  • fDate
    10-16 Nov. 2002
  • Firstpage
    1424
  • Abstract
    Liquid Xe has a fast response (decay time of scintillation light: 2.2 ns, 22 ns and that of recombination light: 45 ns), and high scintillation yield which is comparable to that of NaI(Tl) crystal. These properties are considered to be suitable for position emission tomography (PET). We constructed a new PET detector using scintillation light from liquid Xe and measured energy resolution, position resolution and time resolution with plastic scintillator. The energy resolution, the position resolution and the time resolution were 30.0 %, 5.4 mm and 706 ps with one PMT, respectively.
  • Keywords
    liquid scintillation detectors; position sensitive particle detectors; positron emission tomography; sodium compounds; thallium; xenon; 2.2 ns; 22 ns; 45 ns; NaI(Tl) crystal; NaI:Tl; PET; Xe; energy resolution; liquid Xe scintillation detector; plastic scintillator; position emission tomography; position resolution; time resolution; Energy measurement; Energy resolution; Plastics; Position measurement; Positron emission tomography; Scintillation counters; Solid scintillation detectors; Testing; Time factors; Time measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2002 IEEE
  • Print_ISBN
    0-7803-7636-6
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2002.1239588
  • Filename
    1239588