First test of liquid Xe scintillation detector for PET
Author
Nishikido, F. ; Doke, T. ; Kikuchi, J. ; Yamamoto, M. ; Mori, T. ; Takizawa, K.
Author_Institution
Waseda Univ., Tokyo, Japan
Volume
3
fYear
2002
fDate
10-16 Nov. 2002
Firstpage
1424
Abstract
Liquid Xe has a fast response (decay time of scintillation light: 2.2 ns, 22 ns and that of recombination light: 45 ns), and high scintillation yield which is comparable to that of NaI(Tl) crystal. These properties are considered to be suitable for position emission tomography (PET). We constructed a new PET detector using scintillation light from liquid Xe and measured energy resolution, position resolution and time resolution with plastic scintillator. The energy resolution, the position resolution and the time resolution were 30.0 %, 5.4 mm and 706 ps with one PMT, respectively.
Keywords
liquid scintillation detectors; position sensitive particle detectors; positron emission tomography; sodium compounds; thallium; xenon; 2.2 ns; 22 ns; 45 ns; NaI(Tl) crystal; NaI:Tl; PET; Xe; energy resolution; liquid Xe scintillation detector; plastic scintillator; position emission tomography; position resolution; time resolution; Energy measurement; Energy resolution; Plastics; Position measurement; Positron emission tomography; Scintillation counters; Solid scintillation detectors; Testing; Time factors; Time measurement;