DocumentCode
2205983
Title
The influence by temperature and product density to accuracy of magnetostrictive level gauge
Author
Ren, Kaichun
Author_Institution
Dept. of Power Eng., Chongqing Commun. Inst., China
fYear
2004
fDate
21-25 June 2004
Firstpage
248
Lastpage
251
Abstract
Magnetostrictive level gauge is provided with some advantages such as long work-life, high accuracy, easy installation and a little recalibration, which makes it applied more widely. But its measurement accuracy is affected by temperature and product density as its floats are the sensing elements of liquid level and water-oil interface, in which the error caused sometime goes beyond the gauge´s absolute accuracy. The effective measures to reduce the influence, such as decreasing the float´s outer diameter and temperature-compensating are proposed.
Keywords
compensation; computerised instrumentation; magnetostrictive devices; magnetostrictive level gauge; product density; temperature density; temperature-compensating; Density measurement; Electromagnetic waveguides; Fuel storage; Liquid waveguides; Magnetic heads; Magnetostriction; Measurement errors; Steel; Temperature; Transmitters;
fLanguage
English
Publisher
ieee
Conference_Titel
Information Acquisition, 2004. Proceedings. International Conference on
Print_ISBN
0-7803-8629-9
Type
conf
DOI
10.1109/ICIA.2004.1373362
Filename
1373362