DocumentCode :
2209176
Title :
Nerve Signal Monitoring Using an Implantable Microelectrode
Author :
Kim, Yong-Ho ; Kim, Yong-Jun ; Kang-Min Ami ; Chungkeun Lee ; Lee, Jong-Ho ; Lee, Myoungho
Author_Institution :
Yonsei Univ., Seoul
fYear :
2007
fDate :
28-31 Oct. 2007
Firstpage :
1140
Lastpage :
1143
Abstract :
This paper reports an implantable microelectrode for monitoring compound nerve action potential of a peripheral nerve that is regenerated form a severe injury, dissection in this case. The microelectrode is realized using a biocompatible polymer substrate (PI2611, HDMicrosystems) and the gold electrode. It may provide a new method for inspecting functions of a regenerated peripheral nerve while a current method is gait analysis that is a foot-print study. The microelectrode was separated from a carrier wafer by chemical ´etch-release´ method inducing no mechanical stress on the device. The device separation method is advantageous for finer features and fabrication reliability because it can eliminate deformation of small features due to the mechanical stress. The proposed implantable had implanted into a sciatic nerve of a rat for 3 months, and compound action potential from the regenerated nerve was recorded through the microelectrodes.
Keywords :
microelectrodes; prosthetics; biocompatible polymer substrate; chemical etch-release method; gold electrode; implantable microelectrode; nerve signal monitoring; regenerated peripheral nerve; Chemicals; Electrodes; Etching; Fabrication; Gold; Injuries; Microelectrodes; Monitoring; Polymers; Stress;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Sensors, 2007 IEEE
Conference_Location :
Atlanta, GA
ISSN :
1930-0395
Print_ISBN :
978-1-4244-1261-7
Electronic_ISBN :
1930-0395
Type :
conf
DOI :
10.1109/ICSENS.2007.4388608
Filename :
4388608
Link To Document :
بازگشت