DocumentCode :
2211713
Title :
Design of a variable field prototype PET camera
Author :
Wong, Wai-Hoi ; Uribe, Jorge ; Lu, Weizhao ; Hu, Guoji ; Hicks, Keri
Author_Institution :
M.D. Anderson Cancer Center, Texas Univ., Houston, TX, USA
Volume :
2
fYear :
1995
fDate :
21-28 Oct 1995
Firstpage :
972
Abstract :
A prototype PET camera has been designed and is being constructed to test the concept, and develop the engineering design and production methodology for a variable field PET camera. The long term goal of the design is to develop a lower cost, high resolution PET camera. The camera has eight detector heads which form a closely packed octagon detector ring with an average diameter of 44 cm for brain/breast and animal model imaging. The heads can be translated radially to a maximum ring diameter of 70 cm for whole body imaging. In the larger diameter modes, the camera rotates 45° during imaging. The camera heads can be set to intermediate positions to fit the camera to the subject size to maximize detection sensitivity and sampling uniformity. The detector design implemented is the quadrant sharing photomultiplier (PMT) design using circular 19 mm PMT. The camera images 27 slices simultaneously. The BGO detector pitch size is 27×2.7 mm. Preliminary simulation studies have been performed to evaluate the resolution, sensitivity, and sampling uniformity
Keywords :
biomedical equipment; brain; cameras; image resolution; photomultipliers; positron emission tomography; 19 mm; 44 cm; animal model imaging; brain imaging; breast imaging; closely packed octagon detector ring; detection sensitivity maximization; lower cost high resolution PET camera; medical diagnostic imaging; medical instrumentation; nuclear medicine; quadrant sharing photomultiplier design; sampling uniformity; sensitivity; variable field prototype PET camera design; Cameras; Design engineering; Design methodology; Detectors; Head; Image sampling; Positron emission tomography; Production; Prototypes; Testing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference Record, 1995., 1995 IEEE
Conference_Location :
San Francisco, CA
Print_ISBN :
0-7803-3180-X
Type :
conf
DOI :
10.1109/NSSMIC.1995.510429
Filename :
510429