Title :
New digital processor for SPECT camera-correction of multiple-energy photon spatial distortion
Author :
Jiang, J. ; Htun, M. ; Kline, B. ; Lim, C. ; Rappoport, V.
Author_Institution :
Trionix Res. Lab. Inc., Twinsburg, OH, USA
Abstract :
Spatial distortion of a SPECT image is energy dependent, due to the discrete nature of the photomultiplier tube (PMT) array and non-uniform response of a PMT to the event location and different depth of interaction in the crystal with different photon energy. A spatial distortion correction without consideration of different photon energies can not produce a satisfactory result for different type tracers used in imaging, especially multiple radionuclide imaging. A new digital processor developed by Trionix makes the correction of spatial distortion independent of energy. The new processor digitizes and computes the photon location distortion according to its energy, and the correction is automatically applied to on-line data acquisition. The correction of multiple-energy photon spatial distortion effectively improves the accuracy of the camera detecting system as well as the camera imaging capability with an expanded photon energy range from 50 to 560 keV. The multiple window spatial registration of the detector at the edges of useful field of view is improved from 2.5 mm to 0.4 mm by using a Ga-67 source. No obvious difference can be observed using Tc-99m, In-111 or Ga-67 isotopes, as judged from the images of the four section resolution bar and the Jaszczak phantom
Keywords :
biomedical equipment; cameras; medical image processing; photomultipliers; single photon emission computed tomography; 50 to 560 keV; Ga; Ga-67; In; In-111; Jaszczak phantom; SPECT camera; Tc; Tc-99m; Trionix; camera detecting system accuracy; camera imaging capability; depth of interaction; digital processor; four section resolution bar; medical diagnostic imaging; multiple window spatial registration; multiple-energy photon spatial distortion correction; nonuniform response; nuclear medicine; photomultiplier tube array; photon energy; photon location distortion; spatial distortion correction; Cameras; Data acquisition; Detectors; Image edge detection; Isotopes; Optical computing; Photomultipliers; Photonic crystals; Single photon emission computed tomography; Spatial resolution;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference Record, 1995., 1995 IEEE
Conference_Location :
San Francisco, CA
Print_ISBN :
0-7803-3180-X
DOI :
10.1109/NSSMIC.1995.510452