• DocumentCode
    2212872
  • Title

    Noise in ML-EM reconstruction of SPECT images with attenuation correction

  • Author

    Liew, S.C. ; Ng, Y.K. ; Hasegawa, B.H.

  • Author_Institution
    Dept. of Phys., Nat. Univ. of Singapore, Singapore
  • Volume
    2
  • fYear
    1995
  • fDate
    21-28 Oct 1995
  • Firstpage
    1198
  • Abstract
    The authors derive a closed form expression for the pixel variance in the iterative ML-EM reconstruction of a uniform circular source embedded in a uniformly attenuating circular object. The derivation includes both the photon noise in the projection data and the noise in the attenuation map used for attenuation correction. The variance is found to increase with the number of iterations, but converges to a finite value. The noise to signal ratio at convergence is the sum of two terms which correspond to the contributions from the photon noise in the projection data and the attenuation map noise. The noise component due to the photon noise varies inversely with the total number of detected counts in the projection data. The relative variance at infinite iterations is similar in magnitude to that obtained for the filtered backprojection algorithm. The component due to attenuation map noise is proportional to the relative variance of the attenuation coefficient in the attenuation map. These observations are consistent with the results of the authors´ previous computer simulation and experimental studies
  • Keywords
    gamma-ray absorption; image reconstruction; iterative methods; medical image processing; noise; single photon emission computed tomography; attenuation map noise; convergence; detected counts; infinite iterations; medical diagnostic imaging; noise to signal ratio; nuclear medicine; pixel variance; projection data photon noise; uniform circular source; uniformly attenuating circular object; Attenuation; Computer simulation; Embedded computing; Image reconstruction; Iterative algorithms; Physics computing; Radiology; Signal to noise ratio; Single photon emission computed tomography; Uncertainty;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference Record, 1995., 1995 IEEE
  • Conference_Location
    San Francisco, CA
  • Print_ISBN
    0-7803-3180-X
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1995.510476
  • Filename
    510476