2222620
Title
MW energy irradiator for reflexotherapy
Author
Belavskaya, S.V. ; Berezhinskii, I.V. ; Vityugov, V.A. ; Lisitsyna, L.I. ; Likhanov, Y.M.
Author_Institution
Novosibirsk State Tech. Univ., Russia
fYear
2001
fDate
2001
Firstpage
233
Lastpage
235
Abstract
The efficiency of MW power transmission from the end of a coaxial waveguide to a biologically active point (BAP) is considered. The distribution of the power in biological tissue (BT) irradiated by UHF electromagnetic radiation when the end of the coaxial irradiator is in close contact with the surface of the BAP is measured. It is shown that under such conditions 50% of the irradiated power is absorbed in a BT layer of thickness 0.5 mm and 90% in a 1 mm layer
Keywords
biomedical equipment; biothermics; coaxial waveguides; radiation therapy; radiofrequency heating; waveguide components; 0.5 to 1 mm; 50 to 90 percent; 800 to 1200 MHz; 915 MHz; MW energy irradiator; MW power transmission; UHF electromagnetic radiation; biological tissue; biologically active point; coaxial irradiator; coaxial waveguide; power distribution; reflexotherapy; Antenna measurements; Biological tissues; Coaxial components; Dielectric measurements; Electromagnetic radiation; Electromagnetic waveguides; Frequency; Muscles; Personnel; Power measurement;