Abstract :
Experimental and model studies were performed to measure the surface Laplacian using a rectangular finite difference approximation. The experimental approach used 10 normal subjects with two sites on the torso. Electrode spacing was 2 cm. The surface Laplacian is theoretically independent of rotation of the electrode array. The data showed considerable variation with rotation. Model studies employed a realistic 23 dipole source. A spherical volume conductor showed invariance with rotation, as anticipated theoretically. A realistic torso, however, showed variation with rotation, although a not as severe as that measured. The average signal to noise ratio was 3.3 and 2.5 at the two sites. These results raise serious questions about the practical ability to measure the surface Laplacian on the torso
Keywords :
bioelectric potentials; biomedical measurement; finite difference methods; skin; surface potential; 2 cm; accurate recording feasibility; electrode array rotation; electrode spacing; medical measurements; realistic 23 dipole source; signal to noise ratio; skin potentials measurement; surface Laplacian; torso; Biomedical engineering; Conductors; Electrodes; Filters; Finite difference methods; Laplace equations; Performance evaluation; Rotation measurement; Signal to noise ratio; Torso;