    Clinical evaluation of a bedside ST-segment monitor

  • Author

    Burns, Michael P. ; Downs, W. Gregory

  • Author_Institution
    Univ. Hospitals of Cleveland, OH, USA
  • fYear
    1988
  • fDate
    25-28 Sep 1988
  • Firstpage
    97
  • Lastpage
    100
  • Abstract
    The performance of a bedside ST-segment monitor was evaluated on patients in a coronary care unit. The monitor assesses ST segments using dominant normal beats as determined by the monitor´s arrhythmia-detection software. Measurements of ST-segment level are averaged over 1-min intervals, stored, and tended graphically. Control ST-segment measurements were made from magnetic-tape recordings of the ECG. The mean difference between the bedside-monitor and control ST measurements was -0.01 mm, with a standard deviation of ±0.29 mm and a correlation coefficient of 0.90. The monitor accurately trended the ST-segment deviation on a variety of patients over a range of deviation from -2.77 to +3.21 mm
  • Keywords
    biomedical equipment; electrocardiography; patient monitoring; 1 min; bedside ST-segment monitor; clinical evaluation; coronary care unit patients; dominant normal beats; Cardiology; Computerized monitoring; Electrocardiography; Hospitals; Ischemic pain; Magnetic recording; Manufacturing; Measurement standards; Myocardium; Patient monitoring;