DocumentCode :
2232502
Title :
In vivo ultrasound biomicroscopy
Author :
Passmann, C. ; Ermert, Helmut ; Auer, T. ; Kaspar, K.
Author_Institution :
Inst. fur Hochfrequenztech., Ruhr-Univ., Bochum
fYear :
1993
fDate :
31 Oct-3 Nov 1993
Firstpage :
1015
Abstract :
A very high frequency ultrasonic biomicroscopy system (30-130 MHz) using a B-scan combined with a new Z-scan technique for in vivo imaging of the skin has been developed. To get high lateral resolution and a good signal to noise ratio, we utilized a very narrow focused single element ceramic transducer as used in conventional acoustic microscopy resulting in resolutions of 30 μm lateral and 6 μm axial. Due to the transducer beam shape corresponding to a very short usable focal zone, the B-scan technique has been modified by an additional Z-scan in order to get sufficient depth range suitable for dermatologic diagnostics. We proved the resolution capabilities of the system by in vivo imaging experiments of diseased and normal skin
Keywords :
biomedical ultrasonics; 30 mum; 30 to 130 MHz; 6 mum; B-scan; Z-scan; high frequency ultrasonic biomicroscopy; in vivo; lateral resolution; resolution; signal to noise ratio; skin; Acoustic imaging; Acoustic transducers; Focusing; Frequency; High-resolution imaging; In vivo; Signal resolution; Signal to noise ratio; Skin; Ultrasonic imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 1993. Proceedings., IEEE 1993
Conference_Location :
Baltimore, MD
Print_ISBN :
0-7803-2012-3
Type :
conf
DOI :
10.1109/ULTSYM.1993.339636