• DocumentCode
    2247896
  • Title

    Development of high performance immunoassay system using multichannel microchip

  • Author

    Yamanaka, M. ; Sato, K. ; Tokeshi, M. ; Katou, H. ; Kimura, H. ; Kitamori, T.

  • Author_Institution
    Sch. of Eng., Univ. of Tokyo, Japan
  • fYear
    2001
  • fDate
    Oct. 31 2001-Nov. 2 2001
  • Firstpage
    26
  • Lastpage
    27
  • Abstract
    Immunoassay is known as one of the most important analytical methods and widely used in clinical diagnoses and biochemical studies. The conventional immunoassay, however, requires a long assay time. We developed a bead-bed immunoassay system integrated on a microchip which successfully reduces the assay time from 2 days to 30 minutes using simple operations for assay (Sato et al, 2000 and 2001). Here, aiming at high throughput assay, an integrated immunoassay system using a multichannel microchip was developed for simultaneous analyses of multiple samples.
  • Keywords
    biosensors; microfluidics; micromachining; microsensors; microvalves; 2 day; 30 min; analytical methods; assay operations; assay throughput; assay time; biochemical studies; clinical diagnoses; immunoassay; immunoassay system; microchip integrated bead-bed immunoassay system; multichannel microchip; multichannel microchip integrated immunoassay system; simultaneous multiple sample analysis; Calibration; Chemistry; Dry etching; Fabrication; Glass; Immune system; Medical diagnostic imaging; Pressure control; Valves; Wet etching;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microprocesses and Nanotechnology Conference, 2001 International
  • Conference_Location
    Shimane, Japan
  • Print_ISBN
    4-89114-017-8
  • Type

    conf

  • DOI
    10.1109/IMNC.2001.984049
  • Filename