DocumentCode :
2255211
Title :
Safety and performance compliance test of an infant incubator
Author :
Ibrahim, Fatimah ; Ding, Jeannie Wong Hsiu ; Taib, Mohd Nasir ; Babu, P. Satheesh
Author_Institution :
Dept. of Biomed. Eng., Malaya Univ., Kuala Lumpur, Malaysia
fYear :
2002
fDate :
2002
Firstpage :
35
Lastpage :
39
Abstract :
This paper describes the various aspects of safety and performance of a specific commercial infant incubator and its compliance with the set requirements and regulations according to IEC 601-1 and IEC 601-2-19. The qualitative and the quantitative tests of electrical safety have been carried out. In the electrical test, the grounding resistance, leakage current, line voltage, temperature control, and the safety of thermostat were tested for their compliance with the IEC 601/BS5724. The methodology and equipment employed for the compliance tests are also discussed.
Keywords :
IEC standards; biomedical equipment; conformance testing; safety; temperature control; thermostats; IEC standards; grounding resistance; infant incubator; leakage current; line voltage; performance compliance test; quantitative tests; safety aspects; temperature control; thermostat; Biomedical engineering; Circuit testing; Electrical safety; Fasteners; Grounding; IEC standards; Inspection; Pediatrics; Safety devices; Thermostats;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Research and Development, 2002. SCOReD 2002. Student Conference on
Print_ISBN :
0-7803-7565-3
Type :
conf
DOI :
10.1109/SCORED.2002.1033049
Filename :
1033049
Link To Document :
بازگشت