• DocumentCode
    2276503
  • Title

    Capacitance calibration based on two ac quantum hall resistances

  • Author

    Schurr, J. ; Bürkel, V. ; Kibble, B.P.

  • Author_Institution
    Phys.-Tech. Bundesanstalt, Braunschweig
  • fYear
    2008
  • fDate
    8-13 June 2008
  • Firstpage
    108
  • Lastpage
    109
  • Abstract
    We report on the precise calibration of capacitance standards based on two independent ac quantum Hall resistances. The capacitance is thereby traced to a quantum standard of impedance. The advantages of the realized chain and experimental tests of the coaxial ac bridges involved will be discussed.
  • Keywords
    calibration; capacitance measurement; measurement standards; quantum Hall effect; ac quantum hall resistances; capacitance calibration; coaxial ac bridges; impedance; quantum standard; Bridge circuits; Calibration; Coaxial components; Crosstalk; Frequency; Impedance; Measurement standards; Quantum capacitance; Testing; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
  • Conference_Location
    Broomfield, CO
  • Print_ISBN
    978-1-4244-2399-6
  • Electronic_ISBN
    978-1-4244-2400-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2008.4574676
  • Filename
    4574676