Abstract :
This paper presents the acoustical design, fabrication and evaluation of a 96-element 3.2 MHz phased-array ultrasound transducer for echocardiography probe utilizing Pb(Mg1/3Nb2/3)O3-30%PbTiO3 (PMNPT) single crystal. The transducer consisted of a 1-3 PMNPT/epoxy composite with two quarter wave matching layers and a backing. The pulse-echo performance and echo amplitude of our prototypes are found to be superior to a similarly constructed lead zirconate titanate (PZT) array. The increased transducer bandwidth offers significant performance advantages, particularly in penetration and resolution.
Keywords :
biomedical transducers; composite materials; echocardiography; lead compounds; ultrasonic transducer arrays; PMNPT single crystal; PMNPT-epoxy composite; PbMg0.33Nb0.67O3-PbTiO3; acoustical design; backing; constructed lead zirconate titanate array; echo amplitude; echocardiography probe; frequency 3.2 MHz; phased-array ultrasonic medical imaging transducer; prototypes; pulse-echo performance; transducer bandwidth; two-quarter wave matching layers; Acoustic transducers; Biomedical imaging; Biomedical transducers; Echocardiography; Fabrication; Niobium; Probes; Prototypes; Ultrasonic imaging; Ultrasonic transducers;