DocumentCode
2278626
Title
Calibration of four-terminal-pair resistor
Author
DongXue, Dai ; Yongshun, Ruan ; Xiaochao, Wang
Author_Institution
Nat. Inst. of Metrol., Beijing
fYear
2008
fDate
8-13 June 2008
Firstpage
332
Lastpage
333
Abstract
Based on the standard capacitor acting as reference standard from 100 kHz to 10 MHz, the measurement principle of four-terminal-pair impedance and the open-short-load (OSL) technique are used to calibrate the frequency characteristic of four-terminal pair resistor. In order to eliminate or reduce the nonlinearity error of measurement instrument, the method can keep measurement range of instrument invariable by selecting different nominal standard capacitor at the different calibration frequency.
Keywords
calibration; capacitors; error analysis; resistors; calibration; four-terminal-pair resistor; frequency 100 kHz to 10 MHz; frequency characteristics; measurement instrument; nonlinearity error reduction; open-short-load technique; standard capacitor; Calibration; Capacitors; Circuits; Electrical resistance measurement; Equations; Frequency measurement; Impedance measurement; Instruments; Measurement standards; Resistors;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
Conference_Location
Broomfield, CO
Print_ISBN
978-1-4244-2399-6
Electronic_ISBN
978-1-4244-2400-9
Type
conf
DOI
10.1109/CPEM.2008.4574788
Filename
4574788
Link To Document