DocumentCode :
2279202
Title :
Research of transfer quantum hall resistance with direct current comparator bridge instrument
Author :
Huang, X.D. ; Cai, J.Z. ; Sun, Y. ; Li, J.D.
Author_Institution :
Beijing Orient Inst. of Meas. & Test, Beijing, China
fYear :
2008
fDate :
8-13 June 2008
Firstpage :
400
Lastpage :
401
Abstract :
The problems of transfer Quantum Hall resistance to 1 Omega resistance standard with direct current comparator bridge (DCC) are introduced. Some effective technique ways are suggested by analyzing the possible causations of the measurement results´ fluctuation. The comparing experiments between DCC and cryogenic current comparator bridge (CCC) have been performed with analyzing the uncertainty of measurement results. The transfer uncertainty from Quantum Hall resistance to 1 Omega standard with DCC is 3times10-8 (k=1), which is enough accurate in practical resistance standard calibration of metrology.
Keywords :
calibration; current comparators; electric resistance measurement; measurement standards; measurement uncertainty; quantum Hall effect; cryogenic current comparator bridge; direct current comparator bridge; measurement uncertainty; quantum Hall resistance; resistance 1 ohm; resistance standard calibration; transfer uncertainty; Bridges; Calibration; Cryogenics; Current measurement; Electrical resistance measurement; Fluctuations; Instruments; Measurement uncertainty; Performance analysis; Performance evaluation;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
Conference_Location :
Broomfield, CO
Print_ISBN :
978-1-4244-2399-6
Electronic_ISBN :
978-1-4244-2400-9
Type :
conf
DOI :
10.1109/CPEM.2008.4574822
Filename :
4574822
Link To Document :
بازگشت