A multistage technique for current transformers with current ratios of less than unity
Author
So, E. ; Bennett, D.
Author_Institution
Inst. for Nat. Meas. Stand., Nat. Res. Council of Canada, Ottawa, ON
fYear
2008
fDate
8-13 June 2008
Firstpage
464
Lastpage
465
Abstract
An overview of the development and performance of a current transformer with current ratios of less than unity at the National Research Council of Canada (NRC) is described. A prototype current transformer with a current ratio of 0.025A/1A that can be operated at up to at least ten times the rated current (secondary current up to ges 10 A), with ratio errors of less than 20times10-6 in both magnitude and phase with burdens up to 0.1 Omega and at frequencies up to 1000 Hz has been developed.