Abstract
For the accurate measurement and calibration of the famous Tinsley 5685 AC standard resistors, a new type of Kelvin resistance bridge working over frequency range from DC to 100 kHz is described. At the same time, a bridge substitution technique is used in which the unknown Tinsley resistor is compared to the standard of AC/DC calculable resistor with double helix arrangement. Through this bridge the frequency characteristics of these portable AC standard resistors with the resistance value from 1 Omega to 10 kOmega have been obtained, which will extend their frequency range in metrology calibration.
Keywords
bridge circuits; calibration; electric resistance measurement; measurement standards; resistors; Kelvin resistance bridge; Tinsley 5685 AC standard resistors; calibration; frequency 100 kHz; resistance 1 kohm to 10 kohm; Arm; Battery charge measurement; Bridge circuits; Detectors; Electrical resistance measurement; Frequency measurement; Kelvin; Measurement standards; Metrology; Resistors;