DocumentCode :
2280385
Title :
A new method to measure the dielectric loss of Teflon specimen in cylindrical tube shape
Author :
Li, Z. ; Huang, Lu ; Zhang, Z. ; He, Qing ; Zhao, J. ; Han, B.
Author_Institution :
Nat. Inst. of Metrol., Beijing, China
fYear :
2008
fDate :
8-13 June 2008
Firstpage :
556
Lastpage :
557
Abstract :
The Teflon frame is used in establishing NIM´s AC Resistance Calculable Standard. Comparison with the suspended design such as coaxial, quadrifilar and octofilar types, the frame of resistor gives better stability. However, the dielectric loss of Teflon frame will bring additional ac/dc difference. To measure this dielectric loss precisely, a new measurement structure is provided and the substitute method is carried out to decrease the uncertainty from instrument. The measurement uncertainty of dissipation factor is estimated as several parts in 105.
Keywords :
dielectric loss measurement; measurement standards; measurement uncertainty; polymers; AC Resistance Calculable Standard; Teflon frame; ac/dc difference; cylindrical tube shape; dielectric loss measurement; measurement uncertainty; stability; Coaxial components; Dielectric loss measurement; Dielectric losses; Dielectric measurements; Electrical resistance measurement; Instruments; Loss measurement; Resistors; Shape measurement; Stability;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
Conference_Location :
Broomfield, CO
Print_ISBN :
978-1-4244-2399-6
Electronic_ISBN :
978-1-4244-2400-9
Type :
conf
DOI :
10.1109/CPEM.2008.4574900
Filename :
4574900
Link To Document :
بازگشت