DocumentCode :
2280428
Title :
VNIIM-NRC intercomparison of calibration systems for large value capacitors up to 10 μF at frequencies up to 1 kHz
Author :
Semenov, Yu.P. ; So, E.
Author_Institution :
D.l. Mendeleyev Inst. for Metrol. (VNIIM), St. Petersburg
fYear :
2008
fDate :
8-13 June 2008
Firstpage :
558
Lastpage :
559
Abstract :
A number of international comparisons of calibration systems for capacitors up to 1 nF were conducted in the past at frequencies about 1000 Hz. This intercomparison of large value capacitors calibration systems between VNIIM and NRC was done at frequencies of 50 Hz, 60 Hz, and 1000 Hz. The comparison was implemented with a set of transfer standards, developed by VNIIM, with nominal capacitance values (nF) of 1, 10, 100, 1000, and 10,000. The comparison results, including the transfer standard and calibration systems at VNIIM and NRC, will be presented.
Keywords :
calibration; capacitors; VNIIM-NRC intercomparison; calibration systems; capacitance 1 nF; capacitance 10 nF; capacitance 100 nF; capacitance 1000 nF; capacitance 10000 nF; capacitors; frequency 1000 Hz; frequency 50 Hz; frequency 60 Hz; transfer standards; Bridge circuits; Calibration; Capacitance; Capacitors; Electrical resistance measurement; Frequency; Measurement standards; Ovens; Resistance heating; Standards development;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
Conference_Location :
Broomfield, CO
Print_ISBN :
978-1-4244-2399-6
Electronic_ISBN :
978-1-4244-2400-9
Type :
conf
DOI :
10.1109/CPEM.2008.4574901
Filename :
4574901
Link To Document :
بازگشت