DocumentCode :
2280522
Title :
A PC based setup to determine temperature coefficient of standard inductors using difference voltage measurement technique
Author :
Sharma, Shally ; Saxena, A.K. ; Saleem, Mohd
Author_Institution :
Calibration Labs., Nat. Board of Accreditationfor Testing, New Delhi
fYear :
2008
fDate :
8-13 June 2008
Firstpage :
570
Lastpage :
571
Abstract :
This paper describes a setup to determine the temperature coefficient of standards inductors. Proposed method is based on Transformer Ratio Bridge and difference voltage measurement technique. The temperature coefficient of three inductors of value 1 mH (GR 1482) was measured by this method in the temperature range of 22degC to 27degC, which lies between 42 to 47 ppm/ degC with deviation of plusmn 5 ppm. This method also gives the time taken by inductors to get stabilized against change in temperature.
Keywords :
inductors; temperature measurement; voltage measurement; PC based setup; difference voltage measurement; standards inductors; temperature coefficient; transformer ratio bridge; Bridge circuits; Circuit testing; Electrical resistance measurement; Inductance; Inductors; Laboratories; Measurement standards; Temperature distribution; Temperature measurement; Voltage measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
Conference_Location :
Broomfield, CO
Print_ISBN :
978-1-4244-2399-6
Electronic_ISBN :
978-1-4244-2400-9
Type :
conf
DOI :
10.1109/CPEM.2008.4574907
Filename :
4574907
Link To Document :
بازگشت