  • Abstract
    In December 2007, a comparison of the 10 V Josephson Array Voltage Standard (JAVS) of the LNE was performed with that of the BIPM in the framework of BIPM.EM-KlO.b comparison. During the week allotted to the exercise, the measurement setup was optimized to achieve a relative voltage difference of 7times10-12 within a relative Type A uncertainty of 8times10-12. We then took the opportunity to carry out a direct comparison of a 10 V PTB programmable array against the LNE JAVS. We report on the experiments performed to establish the uncertainty budget. We present the characterization of a 10 V SINIS array. We conclude with the results of a direct comparison against the LNE JAVS.
  • Keywords
    Josephson effect; measurement standards; measurement uncertainty; voltage measurement; BIPM; Josephson array voltage standard; LNE JAVS; PTB programmable array; SIS array; measurement uncertainty; optimized measurement setup; programmable SINIS array; relative type A uncertainty; relative voltage difference; voltage 10 V; Data acquisition; Detectors; Electrical resistance measurement; Frequency estimation; Measurement standards; Microwave frequencies; Sensor arrays; Stability; Uncertainty; Voltage measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
  • Conference_Location
    Broomfield, CO
  • Print_ISBN
    978-1-4244-2399-6
  • Electronic_ISBN
    978-1-4244-2400-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2008.4574913
  • Filename
    4574913