DocumentCode :
2280811
Title :
An improved automated calibration system for 10V DC reference
Author :
Li, H. ; Gao, Y.
Author_Institution :
Nat. Inst. of Metrol., Beijing
fYear :
2008
fDate :
8-13 June 2008
Firstpage :
604
Lastpage :
605
Abstract :
This paper describes an DC voltage calibration system which is improved to be more accurate and more easy to operate. It is based on 10 V Josephson array, and uses a fully automated system consisting of switches setup well designed for accurate measurement. An overall standard uncertainty is estimated less than 5 nV(1sigma) at 10 V level.
Keywords :
Josephson effect; calibration; measurement standards; measurement uncertainty; superconducting arrays; voltage measurement; DC reference; DC voltage calibration system; Josephson array; automated calibration system; measurement uncertainty; voltage 10 V; Automatic control; Calibration; Control systems; Diodes; Engines; Metrology; Phased arrays; Switches; Uncertainty; Voltage;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
Conference_Location :
Broomfield, CO
Print_ISBN :
978-1-4244-2399-6
Electronic_ISBN :
978-1-4244-2400-9
Type :
conf
DOI :
10.1109/CPEM.2008.4574924
Filename :
4574924
Link To Document :
بازگشت