• DocumentCode
    2281510
  • Title

    Electromagnetic compatibility in platinum resistance thermometers

  • Author

    Valencia-Rodriguez, J. ; Figueroa, J.M.

  • Author_Institution
    Centro de Investigacion en Ciencia Aplicada y Tecnologia Avanzada-IPN, Mexico City, Mexico
  • fYear
    2002
  • fDate
    16-21 June 2002
  • Firstpage
    84
  • Lastpage
    85
  • Abstract
    The most precise measurement technique for temperature measurements involves platinum resistance thermometers. In order to perform low uncertainty measurements, it is of paramount importance to ensure the electromagnetic compatibility of such thermometers and the rest of the equipment commonly found in a calibration laboratory (namely furnaces, liquid baths and computers).
  • Keywords
    calibration; electromagnetic compatibility; measurement uncertainty; platinum; resistance thermometers; Pt; calibration; electromagnetic compatibility; measurement uncertainty; platinum resistance thermometer; temperature measurement; Calibration; Electrical resistance measurement; Electromagnetic compatibility; Electromagnetic measurements; Measurement techniques; Measurement uncertainty; Performance evaluation; Platinum; Temperature measurement; Thermal resistance;