  • Title

    An experimental evaluation of a cryogenic noise source by a sliding short method in the frequency range of 8 GHz to 12 GHz

  • Author

    Iida, H. ; Shimada, Y. ; Komiyama, K.

  • Author_Institution
    AIST, Nat. Metrol. Inst. of Japan, Tsukuba
  • fYear
    2008
  • fDate
    8-13 June 2008
  • Firstpage
    700
  • Lastpage
    701
  • Abstract
    A cryogenic noise source (CNS) cooled by liquid nitrogen (LN2) is constructed and evaluated in the frequency range of 8 GHz to 12 GHz (WR-90 waveguide band). A noise temperature of the CNS is changed from the boiling temperature of LN2 by a correction factor (CF) related to loss and temperature distribution of a transmission line (TL). We have investigated the CF by a sliding short method using a total-power radiometer. Comparison results of the noise temperature by the radiometer with two calibrated noise sources proved validity of the CF.
  • Keywords
    cryogenics; noise measurement; radiometry; thermal noise; transmission lines; WR-90 waveguide band; boiling temperature; calibrated noise; correction factor; cryogenic noise source; frequency 8 GHz to 12 GHz; liquid nitrogen; noise temperature; total-power radiometer; transmission line; Acoustic reflection; Cryogenics; Extraterrestrial measurements; Frequency measurement; Radio frequency; Radiometry; Temperature distribution; Temperature sensors; Transmission line measurements; Transmission line theory;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
  • Conference_Location
    Broomfield, CO
  • Print_ISBN
    978-1-4244-2399-6
  • Electronic_ISBN
    978-1-4244-2400-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2008.4574972
  • Filename
    4574972