DocumentCode
2281623
Title
An experimental evaluation of a cryogenic noise source by a sliding short method in the frequency range of 8 GHz to 12 GHz
Author
Iida, H. ; Shimada, Y. ; Komiyama, K.
Author_Institution
AIST, Nat. Metrol. Inst. of Japan, Tsukuba
fYear
2008
fDate
8-13 June 2008
Firstpage
700
Lastpage
701
Abstract
A cryogenic noise source (CNS) cooled by liquid nitrogen (LN2) is constructed and evaluated in the frequency range of 8 GHz to 12 GHz (WR-90 waveguide band). A noise temperature of the CNS is changed from the boiling temperature of LN2 by a correction factor (CF) related to loss and temperature distribution of a transmission line (TL). We have investigated the CF by a sliding short method using a total-power radiometer. Comparison results of the noise temperature by the radiometer with two calibrated noise sources proved validity of the CF.
Keywords
cryogenics; noise measurement; radiometry; thermal noise; transmission lines; WR-90 waveguide band; boiling temperature; calibrated noise; correction factor; cryogenic noise source; frequency 8 GHz to 12 GHz; liquid nitrogen; noise temperature; total-power radiometer; transmission line; Acoustic reflection; Cryogenics; Extraterrestrial measurements; Frequency measurement; Radio frequency; Radiometry; Temperature distribution; Temperature sensors; Transmission line measurements; Transmission line theory;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
Conference_Location
Broomfield, CO
Print_ISBN
978-1-4244-2399-6
Electronic_ISBN
978-1-4244-2400-9
Type
conf
DOI
10.1109/CPEM.2008.4574972
Filename
4574972
Link To Document