Title :
The effects of the environmental conditions of some Zener voltage references
Author :
Nishinaka, H. ; Sakamoto, Y. ; Murayama, Y. ; Iwasa, Akio
Author_Institution :
Nat. Metrol. Inst., Agency of Ind. Sci. & Technol., Ibaraki, Japan
Abstract :
We measured variations of output voltages of four Zener standards (Fluke732B) with change of environmental conditions such as humidity, temperature and pressure. Four Zener standards have negligibly small humidity characteristics. The characteristic of the temperature was the order of 0.01 ppm/K. The characteristic of the pressure was approximately 0.02 ppm/10 hPa.
Keywords :
Zener diodes; transfer standards; voltage measurement; Fluke732B; Zener voltage references; environmental conditions; humidity characteristics; output voltages; pressure; temperature; Atmospheric measurements; Battery charge measurement; Humidity measurement; Laboratories; Measurement standards; Metrology; Pressure measurement; Temperature; Testing; Voltage;
Conference_Titel :
Precision Electromagnetic Measurements, 2002. Conference Digest 2002 Conference on
Conference_Location :
Ottawa, Ontario, Canada
Print_ISBN :
0-7803-7242-5
DOI :
10.1109/CPEM.2002.1034772