DocumentCode :
2283702
Title :
Determination of the frequency dependence of capacitance standards with ceramic dielectrics
Author :
Klionsky, M.D. ; Semenov, Yu.P. ; Moodley, A.
Author_Institution :
Mendeleyev (D.I.) Inst. for Metrol., VNIIM, St. Petersburg, Russia
fYear :
2002
fDate :
16-21 June 2002
Firstpage :
346
Lastpage :
347
Abstract :
The procedure for determination of the frequency dependence (FD) of ceramic capacitance standards characteristic is developed. It is based on application of air capacitance standards with calculable FD, a special summation device and a step calibration method. Measuring instruments are used that allow to compare the capacitance of single capacitors with nominally equal sum capacitance of their combinations. FD of ceramic capacitance standards are determined in the range 100 pF-1 /spl mu/F at frequencies from 50 Hz to 1 MHz. The measurement uncertainty is 0.01-0.1% (k=2).
Keywords :
calibration; capacitance measurement; ceramics; dielectric materials; measurement standards; measurement uncertainty; 100 pF to 1 muF; 50 Hz to 1 MHz; air capacitance standard; capacitance standard; ceramic dielectric; frequency dependence; measurement uncertainty; measuring instrument; step calibration method; summation device; Calibration; Capacitance measurement; Capacitance-voltage characteristics; Capacitors; Ceramics; Dielectric measurements; Frequency dependence; Instruments; Measurement uncertainty; Standards development;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements, 2002. Conference Digest 2002 Conference on
Conference_Location :
Ottawa, Ontario, Canada
Print_ISBN :
0-7803-7242-5
Type :
conf
DOI :
10.1109/CPEM.2002.1034864
Filename :
1034864
Link To Document :
بازگشت