DocumentCode
2284063
Title
DC voltage calibrator based on an array of high-temperature superconductor Josephson junctions
Author
Klushin, A.M. ; Borovitskii, S.I. ; Gelikonova, V.D. ; Komkov, A.V. ; Siegel, M.
Author_Institution
Inst. fur Schichten und Grenzflachen, Forschungszentrum Julich GmbH, Germany
fYear
2002
fDate
16-21 June 2002
Firstpage
392
Lastpage
393
Abstract
We are using a quantum DC voltage generated on an array of high-temperature superconductor (HTS) junctions in the developed voltage calibrator with an output voltage up to 10 V. The described concept and the demonstrated parameters of the main parts of the instrument show that the measurement uncertainty of the output voltage will be a few parts in 10/sup 8/.
Keywords
calibration; high-temperature superconductors; measurement standards; superconducting arrays; superconducting junction devices; voltage measurement; 10 V; DC voltage calibrator; HTS junctions; high-temperature superconductor Josephson junction array; instrument parameters; measurement uncertainty; output voltage; quantum DC voltage; Calibration; DC generators; Diodes; High temperature superconductors; Josephson junctions; Microwave antenna arrays; Resistors; Testing; Voltage; Voltmeters;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements, 2002. Conference Digest 2002 Conference on
Conference_Location
Ottawa, Ontario, Canada
Print_ISBN
0-7803-7242-5
Type
conf
DOI
10.1109/CPEM.2002.1034887
Filename
1034887
Link To Document