DocumentCode :
2285013
Title :
Improvements of measuring system on magnetic levitation system for the determination of magnetic flux quantum
Author :
Shiota, F. ; Morokuma, T.
Author_Institution :
Nat. Inst. of Adv. Ind. Sci. & Technol., Japan
fYear :
2002
fDate :
16-21 June 2002
Firstpage :
514
Lastpage :
515
Abstract :
A development of a video frame synchronized data sampling system is in progress for superconducting magnetic levitation system parameters to correct an error of vertical displacement reading due to attitude change of floating body using real time movement analysis in equilibrium condition.
Keywords :
constants; magnetic flux; magnetic levitation; equilibrium condition; floating body; magnetic flux quantum; measuring system; real-time movement analysis; superconducting magnetic levitation system; video frame synchronized data sampling system; Counting circuits; Error correction; Frequency synchronization; Influenza; Instruction sets; Magnetic flux; Magnetic levitation; Optical interferometry; Sampling methods; Superconducting magnets;
fLanguage :