DocumentCode
2301923
Title
Structural and microfluidic analysis of MEMS based out-of-plane hollow silicon microneedle array for drug delivery
Author
Ashraf, M.W. ; Tayyaba, S. ; Afzulpurkar, N. ; Nisar, A. ; Bohez, Erik L J ; Tuantranont, A.
Author_Institution
Sch. of Eng. & Technol., Asian Inst. of Technol. (AIT), Pathumthani, Thailand
fYear
2010
fDate
21-24 Aug. 2010
Firstpage
258
Lastpage
262
Abstract
Microneedles are gaining popularity because of the ability to deliver the drug through the skin at desire therapeutic range. In this paper, design, structural analysis and fabrication of tapered tip out-of-plane hollow silicon microneedle array for transdermal drug delivery (TDD) applications is first presented. Then computational fluid dynamic (CFD) static analysis is presented to investigate the pressure distribution and velocity distribution of fluid through 5 × 5 microneedle array. The static pressure 10 kPa to 130 kPa was applied for fluidic analysis. Inductively coupled plasma (ICP) etcher machine is used to facilitate the isotropic and anisotropic etching process during the fabrication. Finite element method (FEM) using ANSYS rather than analytical system has been used to perform the simulation. The effect of axial and transverse loads on the microneedles during skin insertion is investigated in the stress analysis. The analysis predicts that the resultant stresses due to applied bending and axial loads are in the safe range. The presented research work provides predicted data to fabricate optimized designs of silicon microneedle array for biomedical applications.
Keywords
bending; bioMEMS; computational fluid dynamics; drug delivery systems; finite element analysis; microchannel flow; micromechanics; needles; pipe flow; silicon; stress analysis; ANSYS finite element method; CFD static analysis; ICP etcher machine; MEMS based silicon microneedle array; Si; TDD; axial load effects; computational fluid dynamics; fluid pressure distribution; fluid velocity distribution; hollow silicon microneedle array; out of plane silicon microneedle array; pressure 10 kPa to 130 kPa; silicon microneedle array design; silicon microneedle array fabrication; silicon microneedle array microfluidic analysis; silicon microneedle array structural analysis; stress analysis; tapered tip silicon microneedle array; transdermal drug delivery; transverse load effects; Arrays; Drug delivery; Etching; Fabrication; Force; Silicon; Skin;
fLanguage
English
Publisher
ieee
Conference_Titel
Automation Science and Engineering (CASE), 2010 IEEE Conference on
Conference_Location
Toronto, ON
Print_ISBN
978-1-4244-5447-1
Type
conf
DOI
10.1109/COASE.2010.5584012
Filename
5584012
Link To Document