Abstract :
Two semi-conductor active cold loads (ACLs) to be used as cold references in spaceborne microwave radiometers have been developed. An X-band frequency was chosen, and the target noise temperature value was in the 50 to 100 K range. The ACLs are characterized in the operating temperature range 0-50degC, and long term stability is assessed. To this end a test bed has been developed. This test bed is actually a stable radiometer, and its design is briefly reviewed. The test setup is described, and preliminary test campaign results indicate output temperatures of 77 K and 55 K for the two ACLs. The temperature sensitivity is slightly below 0.4 K/degC for the units, and excellent long term stability prevails.
Keywords :
calibration; geophysical equipment; radiometers; radiometry; test equipment; X-band frequency; active cold loads; cold references; radiometer calibration; semiconductor ACL measurements; spaceborne microwave radiometers; temperature 0 degC to 50 degC; temperature 50 K to 100 K; test bed development; Calibration; Frequency; Microwave measurements; Microwave radiometry; Space cooling; Stability; System testing; Temperature distribution; Temperature measurement; Temperature sensors; Microwave; active cold load; calibration; radiometer;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Microwave Radiometry and Remote Sensing of the Environment, 2008. MICRORAD 2008
Conference_Location :
Firenze
Print_ISBN :
978-1-4244-1986-9
Electronic_ISBN :
978-1-4244-1987-6
Type :
conf
DOI :
10.1109/MICRAD.2008.4579465
Filename :
4579465
Link To Document :
بازگشت