DocumentCode :
231510
Title :
Research on the sensor calibration of pressure transmitter based on the least square method
Author :
Zhiping Wang ; Aidong Xu ; Kai Wang ; Yan Song ; Bingjun Yan
Author_Institution :
Shenyang Inst. of Autom., Shenyang, China
fYear :
2014
fDate :
19-23 Oct. 2014
Firstpage :
389
Lastpage :
392
Abstract :
The accuracy of capacitance pressure sensor is easily disturbed by work environment. In order to improve the accuracy of pressure sensor, the paper researches on the sensor calibration of pressure transmitter based on the least square method. The measurement circuit is designed to convert the output of sensor from the capacitance value to the cycle value. Working quadric surface is fitted with the value of environmental temperature, the output of pressure sensor and setting pressure by the least square method. The pressure transmitters can accurately measure at different temperatures by the experiment.
Keywords :
calibration; capacitive sensors; least squares approximations; pressure sensors; capacitance pressure sensor; least square method; measurement accuracy; measurement circuit; pressure transmitter; sensor calibration; working quadric surface; Abstracts; Accuracy; Instruments; Pressure measurement; Transmitters; pressure transmitter; sensor calibration; the least square method;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Signal Processing (ICSP), 2014 12th International Conference on
Conference_Location :
Hangzhou
ISSN :
2164-5221
Print_ISBN :
978-1-4799-2188-1
Type :
conf
DOI :
10.1109/ICOSP.2014.7015034
Filename :
7015034
Link To Document :
بازگشت