Abstract
In open-ended coaxial dielectric characterisation techniques, intermediate liquid standards between air and water are necessary for precise dielectric measurements of biological samples. Calibration standards with dielectric permittivities much different from those of the measured samples will produce unreliable results. Different intermediate calibration standards are prepared and measurement results are presented. The aim is to obtain various permittivity values between 10 and 80 by diluting methanol and ethanol with distilled water. This technique can also be used for assessing the purity of alcohols.
Keywords
calibration; microwave measurement; organic compounds; permittivity measurement; Et-OH mixtures; Mt-OH mixtures; air; alcohols; biological samples; calibration standards; dielectric permittivities; distilled water; ethanol; frequency 1 GHz to 15 GHz; intermediate liquid standards; methanol; open-ended coaxial dielectric characterisation techniques; relative permittivity measurements;