DocumentCode
2318177
Title
The muscle fiber direction estimation method by the pseudo-unipolar record
Author
Yoshida, Masaki ; Konishi, Yuto ; Nakamura, Hideo ; Hattori, Takumu ; Oka, Hisao ; Ichihashi, Noriaki
Author_Institution
Fac. Biomed. Eng., Osaka Electro-Commun. Univ., Neyagawa, Japan
fYear
2011
fDate
5-7 Oct. 2011
Firstpage
1
Lastpage
4
Abstract
When a muscle fiber direction and an electrode direction are different, some problems will occur in the surface electromyogram measurement. Therefore, by this study, we propose the surface electromyogram measurement procedure that a muscle fiber direction can estimate from data after measurement. This measurement procedure records differential signals between a reference electrode signal and a signal from all electrodes of the matrix electrode. We can obtain the differential signal between any electrodes by the calculation. We placed a matrix electrode and reference electrode on the biceps brachii muscle and recorded surface electromyogram. By this measurement procedure, we were able to estimate the muscle fiber direction from the data after the measurement. We can also choose two electrodes which do not sandwich the neuromuscular junction and along a muscle fiber direction from a matrix electrode.
Keywords
biomedical electrodes; electromyography; neuromuscular stimulation; bicep brachii muscle; differential signals; matrix electrode; muscle fiber direction estimation method; neuromuscular junction; pseudo-unipolar record; reference electrode signal; surface electromyogram measurement procedure; Biomedical measurements; Electrodes; Head; Junctions; Neuromuscular; Optical fiber devices;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Engineering, 2011 10th International Workshop on
Conference_Location
Kos
Print_ISBN
978-1-4577-0553-3
Type
conf
DOI
10.1109/IWBE.2011.6079039