Title :
Study on stabilized zero-point of electromagnetic flowmeter with rapid excitation
Author :
Saito, OK ; Sakurai, Y. ; Okayama, T.
Author_Institution :
Instrum. Div., Hitachi Ltd., Ibaraki, Japan
Abstract :
This paper describes a designing method to prevent zero-point drift which are susceptible to result when a rapid excitation is employed with electromagnetic flowmeter. It is shown how to prevent adverse effect of magnetic flux differential noise in rapid excitation, the new items of the composite alternated constant current excitation circuit, magnetic circuit configuration, and drift-corrected signal processing. As a result, zero-point drift which were observed ±2% in the maximum sensitivity range of 0.1 m/s have been reduced to ±0.15% or less
Keywords :
electromagnetic devices; electromagnetic interference; exciters; flow measurement; flowmeters; interference suppression; random noise; signal processing; composite alternated constant current excitation circuit; drift-corrected signal processing; electromagnetic flowmeter; industrial measurements; magnetic circuit configuration; magnetic flux differential noise; process measurements; rapid excitation; stabilized zero-point; zero-point drift; Circuits; Coils; DC-DC power converters; Electromagnetic measurements; Frequency; Instruments; Magnetic flux; Magnetic noise; Switches; Voltage control;
Conference_Titel :
Instrumentation and Measurement Technology Conference, 1994. IMTC/94. Conference Proceedings. 10th Anniversary. Advanced Technologies in I & M., 1994 IEEE
Conference_Location :
Hamamatsu
Print_ISBN :
0-7803-1880-3
DOI :
10.1109/IMTC.1994.351981