• DocumentCode
    2340055
  • Title

    A filtered-backprojection algorithm for astigmatic SPECT reconstruction

  • Author

    Wang, Huili ; Jaszczak, Ronald J. ; Coleman, R. Edward

  • Author_Institution
    Duke Univ. Med. Center, Durham, NC, USA
  • Volume
    4
  • fYear
    1994
  • fDate
    30 Oct-5 Nov 1994
  • Firstpage
    1925
  • Abstract
    A filtered backprojection algorithm is presented for astigmatic collimation. The reconstruction algorithm is derived based on the image reconstruction algorithm derived by Feldkamp et al. (see J. Opt. Soc. A, vol. 1, p. 612-19, 1984) for cone beam collimation. The derived algorithm is validated based on Monte Carlo simulated data. Using the astigmatic reconstruction algorithm, the authors demonstrate the capability of astigmatic collimation to balance the trade-off among axial blurring, field-of-view, and collimator sensitivity for selecting a convergent beam collimation. Reconstructed images also demonstrate that astigmatic collimation can provide improved axial blurring with increased collimator sensitivity compared to cone beam collimation
  • Keywords
    Monte Carlo methods; image reconstruction; medical image processing; single photon emission computed tomography; Monte Carlo simulated data; astigmatic SPECT reconstruction; astigmatic collimation; axial blurring; collimator sensitivity; cone beam collimation; convergent beam collimation; field-of-view; filtered-backprojection algorithm; medical diagnostic imaging; nuclear medicine; Biomedical imaging; Cancer; Collimators; Focusing; Geometry; Image converters; Image reconstruction; Monte Carlo methods; Radiology; Reconstruction algorithms;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference, 1994., 1994 IEEE Conference Record
  • Conference_Location
    Norfolk, VA
  • Print_ISBN
    0-7803-2544-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1994.474690
  • Filename
    474690