  • Abstract
    This paper proposes a new wireless biopsy method where a magnetically actuated untethered soft capsule endoscope carries and releases a large number of thermo-sensitive, untethered microgrippers (μ-grippers) at a desired location inside the stomach and retrieves them after they self-fold and grab tissue samples. We describe the working principles and analytical models for the μ-gripper release and retrieval mechanisms, and evaluate the proposed biopsy method in ex vivo experiments. This hierarchical approach combining the advanced navigation skills of centimeter-scaled untethered magnetic capsule endoscopes with highly parallel, autonomous, submillimeter scale tissue sampling μ-grippers offers a multifunctional strategy for gastrointestinal capsule biopsy.
  • Keywords
    biomedical engineering; endoscopes; grippers; medical robotics; gastrointestinal capsule biopsy; magnetic capsule endoscope; retrieval mechanism; stomach; thermosensitive microgrippers; untethered microgrippers; wireless biopsy; Biopsy; Cameras; Endoscopes; Magnetic resonance imaging; Magnetic separation; Robots; Biopsy; capsule endoscopy; medical robotics; microgrippers (μ-grippers); minimally invasive surgery;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2013.2283369
  • Filename
    6607189