• DocumentCode
    2346274
  • Title

    Calibration and correction of a modular cylindrical SPECT system (McSPECT-II)

  • Author

    Jin, Yongjie ; Liu, Jingai ; Chang, Wei ; Xu, Xiangjun ; Liu, Yinong ; Huang, Gang

  • Author_Institution
    Rush Presbyterian-St. Lukes Med. Center, Chicago, IL, USA
  • Volume
    4
  • fYear
    1994
  • fDate
    30 Oct-5 Nov 1994
  • Firstpage
    1777
  • Abstract
    The authors have developed a multi-crystal SPECT (McSPECT-II) system for clinical brain imaging. This system features a stationary cylindrical detector and a rotating collimator. The cylindrical detector is comprised of 120 NaI(Tl) crystal bars coupled to 24 columns ×3 rows of PMTs. Each PMT output is digitized by an ADC. Position calculations, and the associated calibrations and corrections are performed using software and look-up-tables on a Pentium-based PC. The collimator has four quadrants, each consists of a parallel array of collimator units. The four corner wedges between adjacent quadrants are utilized to accommodate line sources for system calibrations. The authors report the calibration and correction procedures developed for the McSPECT-II system. These procedures are: (1) azimuthal bar identification, (2) floating energy windows, (3) axial linearity correction and (4) uniformity correction. Results of these calibrations and corrections are also reported
  • Keywords
    biomedical equipment; brain; calibration; single photon emission computed tomography; McSPECT-II; NaI:Tl; Pentium-based PC; SPECT equipment calibration; SPECT equipment correction; axial linearity correction; azimuthal bar identification; clinical brain imaging; collimator; corner wedges; crystal bars; floating energy windows; look-up-tables; medical diagnostic imaging; medical instrumentation; modular cylindrical SPECT system; multicrystal SPECT; nuclear medicine; parallel array; rotating collimator; stationary cylindrical detector; uniformity correction; Bars; Brain; Calibration; Collimators; Data acquisition; Detectors; Engine cylinders; Linearity; Optical imaging; Photonic crystals;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference, 1994., 1994 IEEE Conference Record
  • Conference_Location
    Norfolk, VA
  • Print_ISBN
    0-7803-2544-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1994.474720
  • Filename
    474720