• DocumentCode
    2349984
  • Title

    6F-2 Elasticity Imaging Using High Frame Rate Ultrasound Imaging

  • Author

    Park, Suhyun ; Aglyamov, Salavat R. ; Scott, W. Guy ; Emelianov, Stanislav Y.

  • Author_Institution
    Dept. of Biomed. Eng., Texas Univ. at Austin, TX
  • fYear
    2006
  • fDate
    2-6 Oct. 2006
  • Firstpage
    602
  • Lastpage
    605
  • Abstract
    We investigated the performance of the conventional and high frame rate ultrasound imaging techniques for strain and elasticity imaging. High frame rate imaging is necessary to track fast deformation in elasticity imaging, but the ultrasound image quality is known to be compromised. Both numerical and experimental studies were performed using point targets and tissue mimicking phantoms. The results of our study confirm that the signal to noise ratio (SNR), contrast to noise ratio (CNR), and axial/lateral resolution of the displacement and strain images acquired using high frame rate ultrasound imaging are comparable with those obtained using the conventional imaging. Thus, the high frame rate ultrasound imaging can be reliably used for both static and dynamic elasticity imaging if the frame rate to capture the internal tissue motion is critical
  • Keywords
    biological tissues; biomedical ultrasonics; elasticity; phantoms; ultrasonic imaging; contrast-to-noise ratio; elasticity imaging; frame rate; signal-to-noise ratio; strain imaging; tissue mimicking phantoms; tissue motion; ultrasound imaging; Capacitive sensors; Elasticity; High-resolution imaging; Image quality; Image resolution; Imaging phantoms; Signal resolution; Signal to noise ratio; Target tracking; Ultrasonic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2006. IEEE
  • Conference_Location
    Vancouver, BC
  • ISSN
    1051-0117
  • Print_ISBN
    1-4244-0201-8
  • Electronic_ISBN
    1051-0117
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2006.157
  • Filename
    4152013