6G-1 Characterization of In Vivo Atherosclerotic Plaques in the Carotid Artery with Acoustic Radiation Force Impulse Imaging
Author
Dahl, Jeremy J. ; Dumont, Douglas M. ; Miller, Elizabeth M. ; Schwark, Earl ; Allen, Jason D. ; Trahey, Gregg E.
Author_Institution
Dept. of Biomed. Eng., Duke Univ., Durham, NC
fYear
2006
fDate
2-6 Oct. 2006
Firstpage
706
Lastpage
709
Abstract
Acoustic Radiation Force Impulse (ARFI) imaging is a useful method for characterizing the local mechanical properties of tissue. ARFI imaging uses high energy, focused ultrasound pulses to generate small displacements (1-10 mum) in tissue. The response of tissue to these localized forces is observed using a series of imaging pulses, which track displacement of the tissue. ARFI imaging is complementary to B-mode imaging, and can readily distinguish between stiff and soft regions of tissue with high contrast. Currently, physicians cannot easily determine whether atherosclerotic plaques in carotid arteries are vulnerable to rupture. We are investigating the use of ARFI imaging to characterize atherosclerotic plaque in carotid arteries and to guide the treatment of these plaques. We have created custom beam sequences on a Siemens Antarestrade scanner to implement ARFI imaging. Five patients with atherosclerotic plaque and four volunteers without known atherosclerotic plaque underwent B-Mode and ARFI scanning of their carotid arteries. Measurements of the complete vessel wall (adventitial layer plus IMT) with ARFI imaging are shown to be consistent with measurements of the intima-media thickness (IMT) in normal B-mode scans. In plaque-filled vessels, the mean near-wall IMT was 0.92 mm and the mean far-wall IMT was 1.24 mm. In ARFI imaging, vessel walls have mean thicknesses of 2.17 mm and 2.17 mm for the near and far walls, respectively. For volunteers without plaques, the near and far wall IMT measurements were 0.52 and 0.54 mm, respectively and the ARFI vessel thicknesses were 1.39 and 1.26 mm. Both IMT measurements and ARFI measurements were significantly (p < 0.005) greater for patients with plaques. We demonstrate high-resolution ARFI images of carotid plaques depicting their complex mechanical structure, which often involve regions of soft and hard plaque. We discuss methods of improving the contrast and resolution of ARFI vascular images
Keywords
biomechanics; biomedical ultrasonics; cardiology; deformation; patient diagnosis; thickness measurement; 0.52 mm; 0.54 mm; 0.92 mm; 1 to 10 micron; 1.24 mm; 1.26 mm; 1.39 mm; 2.17 mm; ARFI carotid artery scanning; ARFI imaging; B-Mode carotid artery scanning; IMT; Siemens Antares scanner; acoustic radiation force impulse imaging; adventitial layer; focused ultrasound pulses; imaging pulses; in vivo atherosclerotic plaques; intima-media thickness; soft tissue regions; stiff tissue regions; tissue displacement tracking; tissue mechanical properties; Acoustic beams; Acoustic imaging; Acoustic pulses; Carotid arteries; Focusing; In vivo; Mechanical factors; Pulse generation; Thickness measurement; Ultrasonic imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 2006. IEEE
Conference_Location
Vancouver, BC
ISSN
1051-0117
Print_ISBN
1-4244-0201-8
Electronic_ISBN
1051-0117
Type
conf
DOI
10.1109/ULTSYM.2006.192
Filename
4152048
Link To Document