• DocumentCode
    2350997
  • Title

    1I-1 Integrated Endoscope for Real-Time 3D Ultrasound Imaging and Hyperthermia: A Feasibility Study

  • Author

    Pua, Eric C. ; Qiu, Yupeng ; Smith, Stephen W.

  • Author_Institution
    Dept. of Biomed. Eng., Duke Univ., Durham, NC
  • fYear
    2006
  • fDate
    2-6 Oct. 2006
  • Firstpage
    837
  • Lastpage
    840
  • Abstract
    The goal of this research is to determine the feasibility of using a single endoscopic probe for the combined purpose of real-time 3D (RT3D) ultrasound imaging of a target organ and the delivery of ultrasound hyperthermia to facilitate the absorption of compounds for cancer treatment. The ability to combine ultrasound hyperthermia and 3D imaging could improve visualization and targeting. In this study, numerical modeling and experimental measurements were developed to determine the feasibility of combined therapy and imaging with a 5 MHz, 1 cm diameter endoscopic RT3D probe with 504 active channels. A finite element mesh was constructed with over 128,000 elements in LS-DYNA to simulate the induced temperature rise from our transducer with a 3 cm deep focus in tissue. Based on intensity values calculated in Field II using the transducer´s array geometry, a minimum ISPTA of 3.43 W/cm2 is required in order to induce a temperature rise of 4 degC within 5 minutes. It was determined through dosimetry measurements that the probe can presently maintain intensity values up to 2.4 W/cm2 over indefinite times for therapeutic applications combined with intermittent 3D scanning to maintain targeting. In vitro heating experiments of excised pork tissue yielded a maximum temperature rise of 2.3 degC over 5 minutes of ultrasound exposure with an average rise of 1.8 plusmn 0.2 degC over 5 trials. Modifications to the power supply and transducer array may enable us to reach the higher intensities required to facilitate drug delivery therapy
  • Keywords
    biomedical transducers; biomedical ultrasonics; endoscopes; hyperthermia; patient treatment; ultrasonic transducer arrays; 1 cm; 5 MHz; 5 mins; Field II; LS-DYNA; cancer treatment; dosimetry; drug delivery therapy; endoscope; endoscopic RT3D probe; finite element mesh; pork tissue; real-time 3D ultrasound imaging; transducer array; ultrasound hyperthermia; Absorption; Cancer; Endoscopes; Hyperthermia; Medical treatment; Probes; Temperature; Transducers; Ultrasonic imaging; Visualization;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2006. IEEE
  • Conference_Location
    Vancouver, BC
  • ISSN
    1051-0117
  • Print_ISBN
    1-4244-0201-8
  • Electronic_ISBN
    1051-0117
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2006.218
  • Filename