DocumentCode :
2352252
Title :
Detailed investigation of transmission and emission data smoothing protocols and their effects on emission images
Author :
Chatziioannou, A. ; Dahlbom, M.
Author_Institution :
Sch. of Med., California Univ., Los Angeles, CA, USA
Volume :
4
fYear :
1994
fDate :
30 Oct-5 Nov 1994
Firstpage :
1568
Abstract :
Measured attenuation correction in PET, is routinely performed using transmission scans. Acquisition time and noise considerations necessitate low pass filtering of the transmission data, before generating the attenuation correction matrix. This smoothing operation reduces noise propagation from transmission into emission data, but also introduces image artifacts which are mostly pronounced around areas of strongly varying attenuation coefficients. The source of these artifacts, which lies in the mismatch of the spatial resolutions of emission and transmission data, was investigated in this study. The effects of different transmission and emission sinogram smoothing protocols on the emission images were also investigated. A method is proposed that addresses the problem, in coordination with the filtering step during reconstruction. Instead of the standard low pass filtering of the emission data during reconstruction, emission and transmission sinograms can be filtered to the desired reconstructed image resolution prior to reconstruction. This operation reduces or eliminates resolution mismatch and consequent image artifacts, which can be significant, especially in cardiac studies. The proposed method improves the accuracy of the activity distribution in the emission images, with minimal computational and SNR cost
Keywords :
cardiology; image reconstruction; image resolution; medical image processing; positron emission tomography; protocols; smoothing methods; PET; acquisition time; attenuation correction; attenuation correction matrix; cardiac studies; emission data smoothing protocols; emission images; filtering step; image artifacts; image resolution; low pass filtering; noise considerations; noise propagation; reconstruction; resolution mismatch; sinogram smoothing protocols; smoothing operation; spatial resolutions; standard low pass filtering; strongly varying attenuation coefficients; transmission data smoothing protocols; transmission scans; Attenuation measurement; Filtering; Image reconstruction; Image resolution; Low pass filters; Noise generators; Noise reduction; Performance evaluation; Positron emission tomography; Smoothing methods;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1994., 1994 IEEE Conference Record
Conference_Location :
Norfolk, VA
Print_ISBN :
0-7803-2544-3
Type :
conf
DOI :
10.1109/NSSMIC.1994.474763
Filename :
474763
Link To Document :
بازگشت