Title :
Development of prosthetic aortic heart valves
Author :
Ulman, S.A. ; Vyas, B.D. ; Puniyani, R.R.
Author_Institution :
Sch. of Biomed. Eng., Indian Inst. of Technol., Bombay, India
Abstract :
Investigates various new prototypes of aortic valves. The fabricated valve prototypes were then tested by various hydrodynamic and hemocompatibility tests. The various prototype values designed and fabricated were: (1) a bileaflet valve, (2) a conical poppet valve, (3) a disc poppet valve, and (4) a fluidic nozzle valve. These valves were tested for percentage regurgitation across them, for pressure drop in steady flow, for flow visualization in steady and pulsatile flows, and for hemocompatibility. The poppet valves had a low percentage of regurgitation, and at higher heart rates the fluidic nozzle diode was very competent. The study and results serve, at best, as guidelines for future research work in the field. The fluidic nozzle diode, by its very form, appears very promising in the mitral area
Keywords :
artificial organs; blood flow measurement; flow visualisation; pulsatile flow; valves; bileaflet valve; conical poppet valve; critical opening pressure; disc poppet valve; flow visualization; fluidic nozzle valve; heart rate; hemocompatibility tests; hydrodynamic tests; mitral area; percentage regurgitation; pressure gradient; prosthetic aortic heart valve prototypes; pulsatile flow; steady flow pressure drop; Diodes; Fluid flow measurement; Heart valves; Mercury (metals); Performance evaluation; Pressure measurement; Prosthetics; Prototypes; Testing; Visualization;
Conference_Titel :
Engineering in Medicine and Biology Society, 1995 and 14th Conference of the Biomedical Engineering Society of India. An International Meeting, Proceedings of the First Regional Conference., IEEE
Conference_Location :
New Delhi
Print_ISBN :
0-7803-2711-X
DOI :
10.1109/RCEMBS.1995.532969