DocumentCode
2355684
Title
P2L-4 Ultrasonic Detection and Quantitative Analysis of Microscopic Bubbles and Particles in Solution: Enhanced by Attenuation Compensation
Author
Brand, Sebastian ; Klaua, Robert ; Dietrich, Georg ; Schultz, Michael
Author_Institution
Dept. of Orthopaedics, Halle-Wittenberg Univ., Halle
fYear
2006
fDate
2-6 Oct. 2006
Firstpage
1840
Lastpage
1843
Abstract
In many technical and medical applications microscopic bubbles and particles play an important role. For a non-invasive and quantitative estimation of the particle characteristics acoustic properties along the path of sound propagation are of major concern. When applying clamp-on probes that can be used with a variety of tubes the compensation of the acoustic attenuation is crucial for reliable parameter estimation. In addition, the acoustic properties of the tube material vary as a function of environmental conditions. An ultrasonic probe that enabled an automated and real-time compensation for attenuation properties was under investigation. The set up employs ultrasound in pulsed wave Doppler mode to provide a higher sensitivity for moving objects. Experiments on different tubes (wall-thickness and material) were performed. Glass beads with Gaussian distributed diameters between 300 mum and 400 mum were used as bubble-phantoms. The means and standard deviations of the particle sizes were estimated from the compensated and uncompensated signals. Size estimates were in the range between 100 and 700 mum with a standard deviation of up to 100% when using uncompensated signals. Standard deviations of the estimates applying the compensation algorithm were 50% with mean values of approximately 380 mum. The described method for attenuation compensation was implemented in the bubble counter system BCC200 (GAMPT mbH, Zappendorf, Germany). It enables a precise monitoring of the occurrence of microscopic bubbles and solid particles for the application in open heart surgery and will increase the patient safety. With its performance the BCC200 can also be employed for the development of arterial filters and oxygenators in biomedical applications and for quality control in industrial applications
Keywords
biomedical ultrasonics; bubbles; parameter estimation; particle size; ultrasonic absorption; ultrasonic measurement; ultrasonic propagation; BCC200; acoustic attenuation; arterial filters; attenuation compensation; biomedical applications; bubble counter system; bubble detection; extracorporal circuit; microscopic bubbles; open heart surgery; oxygenators; parameter estimation; particle characteristics; particle sizes; patient safety; pulsed wave Doppler mode; quality control; solid particles; sound propagation; ultrasonic detection; Acoustic materials; Acoustic propagation; Acoustic signal detection; Attenuation; Biomedical acoustics; Biomedical equipment; Medical services; Microscopy; Parameter estimation; Probes;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 2006. IEEE
Conference_Location
Vancouver, BC
ISSN
1051-0117
Print_ISBN
1-4244-0201-8
Electronic_ISBN
1051-0117
Type
conf
DOI
10.1109/ULTSYM.2006.470