    Design of biomechanically compliant small diameter vascular grafts and their in vitro and in vivo properties

  • Author

    Gupta, B.S. ; Kasyanov, V.A.

  • Author_Institution
    North Carolina State Univ., Raleigh, NC, USA
  • fYear
    1995
  • fDate
    7-9 Apr 1995
  • Firstpage
    212
  • Abstract
    The performances of textile composite vascular grafts, specially designed and constructed in this work were investigated. In developing compliant structures, textile threads of two widely different deformative characteristics, one matching nearly those of the elastin and other of the collagen fibres, were selected. The materials used were a polyurethane monofilament yarn with low modulus of elasticity and a bulked polyester multifilament yarn with high modulus of elasticity. Tubular grafts of diameter in the range 4-6 mm were made by a weaving process utilizing a foil ribbon loom. Two types of woven grafts were developed. In the first type (HVGI1), polyester threads of 9 tex linear density were used as warp, and the same polyester and pre-stretched polyurethane (7.8 tex linear density) were used as the weft. In the second variant (HVG-2), pre-stretched polyurethane thread combined with polyester were used as both the warp and the weft threads. The grafts obtained were stretchable and thus compliant in both the transverse and the longitudinal directions. The hydraulic permeability was assessed at fixed pressure of 120 mmHg. The surface and pore characteristics of grafts were examined using a scanning electron microscope. Eight HVG-I hybrid grafts were also implanted in the carotid and femoral regions of mongrel dogs for periods of up to one year. The form of the pulse wave found in the graft, the healing characteristics of the surface and the changes in the mechanical properties as a result of implantation were examined
  • Keywords
    biomechanics; prosthetics; 1 y; 120 mmHg; 4 to 6 mm; biomechanically compliant small diameter vascular grafts; bulked polyester multifilament yarn; collagen fibres; deformative characteristics; elasticity modulus; elastin; hydraulic permeability; in vitro properties; in vivo properties; mechanical properties changes; mongrel dogs; polyester threads; polyurethane monofilament yarn; pore characteristics; prestretched polyurethane thread; pulse wave; surface characteristics; surface healing characteristics; tubular grafts; vascular grafts design; warp; weft; Biological materials; Dogs; Elasticity; Mechanical factors; Permeability; Scanning electron microscopy; Surface waves; Textile fibers; Weaving; Yarn;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering Conference, 1995., Proceedings of the 1995 Fourteenth Southern
  • Conference_Location
    Shreveport, LA
  • Print_ISBN
    0-7803-2083-2
  • Type

    conf

  • DOI
    10.1109/SBEC.1995.514481
  • Filename
    514481