    Spark discharges for surgical applications

  • Author

    Kolb, Juergen F. ; Scully, Noah ; Gregory, Betsy ; Schoenbach, Karl H.

  • Author_Institution
    Frank Reidy Res. Center for Bioelectrics, Old Dominion Univ., Norfolk, VA, USA
  • fYear
    2010
  • fDate
    23-27 May 2010
  • Firstpage
    258
  • Lastpage
    261
  • Abstract
    We have investigated the potential of spark discharges as a minimally invasive surgical tool. The spark was generated between the tips of two 0.4-mm acupuncture needles, which were inserted into the tissue. The needle tips were facing each other at a distance of 7 mm. Using a Blumlein pulse forming network, high voltage pulses with a nominal duration of 380 ns were applied to these electrodes. The tissue-load was matched to the 100-Ohm impedance of the pulse generator by an equivalent resistor. Depending on tissue type (muscle, liver, fat) and pulse amplitude, several tissue-conditioning pulses had to be applied before breakdown could be achieved. During the application of these pre-breakdown pulses, tissue impedances dropped noticeably. The breakdown voltage for this electrode geometry and these pulse parameters was higher than for air (at atmospheric pressure) but much lower than expected for liquids or solids. The energy that was stored in the system and dissipated in the spark at the highest charging voltage was on the order of 1 J. A visual inspection of the tissue damage showed a highly localized incision, running between the tips of the needle electrodes. Tissue was ablated in a nearly straight and narrow channel of 1-2 mm in diameter. No significant damage to the adjacent tissue was observed. The preliminary histological analysis suggests that the damage is mostly caused by the Shockwave associated with the spark. As such, we believe that the method can be used as a surgical tool, in particular, where a high degree of accuracy is required.
  • Keywords
    biological tissues; electrodes; pulse generators; resistors; shock waves; sparks; surgery; Blumlein pulse forming network; acupuncture needle; breakdown voltage; charging voltage; distance 1 mm to 2 mm; distance 7 mm; electrode geometry; equivalent resistor; minimally invasive surgical tool; prebreakdown pulse; pulse generator; resistance 100 ohm; size 0.4 mm; spark discharge; surgical application; time 380 ns; Discharges; Electrodes; Impedance; Needles; Pulse generation; Sparks;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Power Modulator and High Voltage Conference (IPMHVC), 2010 IEEE International
  • Conference_Location
    Atlanta, GA
  • Print_ISBN
    978-1-4244-7131-7
  • Type

    conf

  • DOI
    10.1109/IPMHVC.2010.5958342
  • Filename
    5958342