Abstract :
Design of mismatched filters using phase-coded excitation pulse for high frequency ultrasound imaging is presented in this paper. Biphase-coded pulses were constructed to excite the transducer. Received signals were compressed with mismatched filters optimized by minimizing peak sidelobe level (PSL). Both simulation and experiments were carried out to demonstrate the advantage of this technique. The simulation results demonstrated a possible sidelobe reduction (<-90 dB) with a slight decrease of the signal to noise ratio of less than 1 dB compared with the compression using matched filters alone. The experimental results showed about 14 dB SNR improvement as well as -40 dB sidelobe level when the Barker-13 code excitation with 3-cycle sinusoidal wave carrier was used
Keywords :
phase coding; ultrasonic imaging; ultrasonic transducers; 3-cycle sinusoidal wave carrier; Barker-13 code excitation; biphase-coded excitation pulse; high frequency ultrasound imaging; mismatched filter; peak sidelobe level; sidelobe reduction; signal to noise ratio; ultrasonic transducer; Biomedical imaging; Biomedical transducers; Decoding; Frequency modulation; Matched filters; Signal to noise ratio; Spatial resolution; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 2006. IEEE
Conference_Location :
Vancouver, BC
ISSN :
1051-0117
Print_ISBN :
1-4244-0201-8
Electronic_ISBN :
1051-0117
Type :
conf
DOI :
10.1109/ULTSYM.2006.560
Filename :
4152416
Link To Document :
بازگشت