Development of the low - cost blood pressure simulator
Author
Wongkamhang, A. ; Thongpance, N. ; Sangworasil, M. ; Roongprasert, K. ; Matsuura, T.
Author_Institution
Dept. of Biomed. Eng., Rangsit Univ., Phathumthani, Thailand
fYear
2014
fDate
26-28 Nov. 2014
Firstpage
1
Lastpage
4
Abstract
The objective of this research was to develop the low - cost blood pressure simulator for verifying of oscillometric automated sphygmomanometer in Thailand and compared this simulator with commercial simulator Fluke Cufflink. The testing results of the precision and the accuracy of pressure generation showed that the average percentage error and percentage precision were 0.5 % and 100 % successively. Finally, oscillometric BP devices (6 models from 5 manufacturers) were tested for checking the performance of the simulator. The results of the performance testing showed that the percentage error and precision were ol % and 99 % sequentially. We expected to continue to improve for the better quality and will be commercialized to be used to verify the automated sphygmomanometer in Thailand afterwards.
Keywords
biomedical engineering; blood pressure measurement; oscillations; simulation; average percentage error; commercial simulator Fluke Cufflink; low-cost blood pressure simulator development; oscillometric BP device; oscillometric automated sphygmomanometer; percentage precision; pressure generation accuracy; Biology; ISO; ISO standards; Performance evaluation; Pressure measurement; Uniform resource locators; Automated Sphygmomanometer; Blood Pressure; Blood Pressure Simulator;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Engineering International Conference (BMEiCON), 2014 7th
Conference_Location
Fukuoka
Type
conf
DOI
10.1109/BMEiCON.2014.7017390
Filename
7017390
Link To Document